Medizin- & Ästhetik-Anlagen

Bei Laserbehandlungen, ob Ablativ oder gütegeschaltet mit Nano- oder Pico-Impulsen, werden Rauchgase und oft virusbelastetes Gewebe freigesetzt.

TBH Filter- und Absauganlagen fangen diese potenziell infektiösen Partikel effektiv auf, um Arzt, Personal und Patienten optimal zu schützen.

Vimeo

By loading the video, you agree to Vimeo's privacy policy.
Learn more

Load video

Testimonial Dr. Hahn

Zur Person: Dr. med. Helmut Hahn ist Dermatologe, Partner im Haut- & Laserzentrum Berchtesgaden und Tagungspräsident des DDL-Kongresses. Im Video erklärt er, warum Berufsgenossenschaften den Einsatz von Absauganlagen zum Schutz vor in der Luft befindlichen, abgetragenen Partikeln fordern.

Welche Medizin-/ Ästhetik- Anlage passt zu mir?

Medizin- und Ästhetik-Anlagen Videos

Vimeo

By loading the video, you agree to Vimeo's privacy policy.
Learn more

Load video

Vimeo

By loading the video, you agree to Vimeo's privacy policy.
Learn more

Load video

Vimeo

By loading the video, you agree to Vimeo's privacy policy.
Learn more

Load video

Vimeo

By loading the video, you agree to Vimeo's privacy policy.
Learn more

Load video

Service & Verkauf

FAQ & Downloads

Nein, die TBH Filter- und Absauganlagen sind keine Medizinprodukte gemäß § 3 Abs. 1 MPG, da sie nicht direkt mit Patienten in Berührung kommen und primär dem Arbeitsschutz dienen. Sie erfüllen nicht die Kriterien für eine medizinische Zulassung nach DIN EN 60601-1. Allerdings können sie zur Erfüllung erhöhter Sicherheitsanforderungen im medizinischen Bereich nach DIN EN 62353 (VDE 0751-1) geprüft werden.

Nein, eine Zulassung ist nicht erforderlich, da die Anlagen so konzipiert sind und der Absaugarm so positioniert ist, dass sie den Patienten während der Anwendung, beim Absaugen, nicht direkt berühren.

TBH Health Basic und TBH Health Pro haben viele Gemeinsamkeiten, wie z.B. ähnliche Luftdurchsatzmengen von bis zu 200m³/h, identische Abmessungen und identische Filtertypen, einschließlich des patentierten InLine-Filters.

Der Hauptunterschied zwischen den beiden Geräten liegt im Motortyp und im statischen Druck. Die TBH Health Basic verfügt über einen Bürstenmotor mit einer garantierten Laufzeit von 600 Betriebsstunden und einer durchschnittlichen Lebensdauer von ca. 600-800 Stunden. Obwohl es technisch möglich ist, nur die Bürsten zu wechseln, kann dies die Lebensdauer des Motors verkürzen, da häufig auch der Kollektor beschädigt wird, was zu einem schnelleren Verschleiß der Bürsten führt. Darüber hinaus bietet der TBH Health Basic einen höheren statischen Druck von 14.000 Pa, was jedoch für Anwendungen mit Absaugarmen irrelevant ist.

Aufgrund der begrenzten Lebensdauer des Motors und der Wartungsanforderungen empfehlen wir die TBH Health Basic für Anwendungen, bei denen die Betriebszeit zwischen 5 und 20 Minuten pro Tag liegt. Die TBH Health Pro für Anwendungen mit einer Betriebszeit von über 20min am Tag.

Die TBH Health Pro und TBH Health Pro Silence sind in Technik und Anwendung ähnlich, haben ähnliche Luftstromraten (max. 200m³/h bzw. max. 300m³/h) durch die gleiche Motortechnologie und sind durch den bürstenlosen Motor und das Antriebssystem wartungsfrei. Die Lebensdauer des Motors beträgt ca. 20.000h – 40.000h, wobei wir eine Garantie von 10.000h geben. Beide Anlagen sind mit dem patentierten InLine-Filter ausgestattet und werden für Anwendungen empfohlen, bei denen die Arbeitszeit mehr als 20 Minuten pro Tag beträgt. Doch es gibt wesentliche Unterschiede. Die TBH Health Pro ist kompakter, hat aber eine kleinere Filterfläche und einen höheren Geräuschpegel (62 dB). Die Anlage verwendet einen kompakten 2-stufigen Filter (HEPA-Partikelfilter H13 + Aktivkohle). Im Gegensatz dazu benötigt die TBH Health Pro Silence etwas mehr Platz, bietet aber eine größere Filterfläche und einen niedrigeren Geräuschpegel (53 dB). Er ermöglicht den getrennten Austausch von HEPA-Partikelfilter H14 und Aktivkohlefilter, was zu geringeren Folgekosten führt.

Die TBH Health Pro ist kompakter und lauter (62 dB), ideal bei begrenztem Platzangebot und weniger Dauerbetrieb. Die TBH Health Pro Silence benötigt mehr Platz, ist leiser (53 dB) und hat geringere Folgekosten durch separat austauschbare Filter. Er eignet sich besser für Dauerbetrieb und lärmsensible Umgebungen. Beide sind für Anwendungen von mehr als 20 Minuten pro Tag ausgelegt.

Die Hauptfilter der TBH Health Serie werden durch einen Differenzdrucksensor überwacht, der anzeigt, wenn sie gesättigt sind. Sollte innerhalb eines Jahres keine Sättigung angezeigt werden, ist ein Wechsel der HEPA-Partikelfilter aus hygienischen Gründen erforderlich. Der Aktivkohlefilter, der nicht durch einen Differenzdrucksensor überwacht wird, sollte mindestens alle 6-12 Monate oder bei Geruchsentwicklung gewechselt werden. Der ebenfalls nicht sensorüberwachte InLine-Filter sollte nach dem TBH-Hygienekonzept täglich, jedoch spätestens wöchentlich, gewechselt werden, um eine Verschleppung von Bakterien und Keimen zu vermeiden.

Der Wartungsaufwand für die Anlage ist minimal. Abgesehen von regelmäßigen Filterwechseln und Reinigungen ist die Anlage weitgehend wartungsfrei. Die Oberflächen und der Absaugarm sollten regelmäßig per Wischdesinfektion gereinigt werden, siehe Schritt für Schritt Videoanleitung unter: werden (siehe Reinigungsvideo unter „Luftabsaugung für medizinische Anwendungen – TBH GmbH“).

Darüber hinaus sollten die Einstellschrauben wöchentlich nachgezogen werden, um ein mögliches Absinken des Absaugarms zu verhindern. Weitere Informationen finden Sie in den Anlagenkatalogen sowie in den Betriebs- und Wartungsanleitungen.

Der patentierte InLine-Filter ist eine Innovation speziell für Anwendungen im medizinischen und ästhetischen Bereich. Dieser Filter, der direkt an der Haube angebracht wird, fängt Partikel, Keime und Bakterien direkt an der Entstehungsstelle ab. Dank des „1-Click-Systems“ kann der InLine-Filter in ca. 5 Sekunden gewechselt werden. Die Haube selbst lässt sich leicht demontieren und reinigen. Der InLine-Filter verhindert durch seine Anordnung an der Spitze, dass der Absaugarm verschmutzt, in dem sich sonst Bakterien und Keime vermehren könnten. Dadurch reduziert sich der Reinigungsaufwand der Anlage erheblich. Darüber hinaus verhindert der InLine-Filter das Herausfallen von Partikeln und eventuell Flüssigkeit aus dem Absaugarm und damit die Übertragung von Keimen zwischen den Patienten. Der Hauptfilter der Anlage wird auf diese Weise geschont und die Standzeit des Filters deutlich verlängert.

 

Die Saugspitze hat einen kleineren Durchmesser. Dadurch wird die Sicht weniger behindert. Sie ist jedoch aus hygienischen Gründen nicht die erste Wahl, da sie den Absaugarm nicht vor Verschmutzung und Keimbildung schützt und den Hauptfilter stärker belastet. Sollte die serienmäßig montierte Haube bei der Positionierung am Patienten stören, kann sie durch Lösen der 4 Schrauben einfach abgenommen werden. Die Funktion des InLine-Filters wird dadurch nicht beeinträchtigt.

Die TBH-Hauptfilter in der Anlage sind einfach zu wechseln, es sind keine besonderen Kenntnisse und Werkzeuge erforderlich. Einfach den Anlagendeckel aufmachen, die Filter rausnehmen und die neuen Filter einsetzen. Eine genaue Videoanleitung finden sie hier: Luftabsaugung für medizinische Anwendungen – TBH GmbH  

Der Wechsel der TBH Health-Anlage Hauptfilter in der Anlage ist einfach und erfordert keine speziellen Kenntnisse oder Werkzeuge. Sie müssen lediglich den Deckel der Anlage öffnen, die alten Filter entfernen und die neuen Filter einsetzen. Eine detaillierte Videoanleitung finden Sie unter dem Link: Luftabsaugung für medizinische Anwendungen – TBH GmbH  

Der Zusammenbau ist sehr einfach und erfordert keine besonderen Kenntnisse. Eine detaillierte Videoanleitung finden Sie unter dem Link: Luftabsaugung für medizinische Anwendungen – TBH GmbH  

Der Absaugarm kann durch Drehen der Schrauben genau positioniert werden. Ein Überdehnen des Absaugarms sollte vermieden werden, da sonst die Anlage kippen kann. Eine detaillierte Videoanleitung finden Sie unter dem Link: Luftabsaugung für medizinische Anwendungen – TBH GmbH  

Das TBH-Hygienekonzept wurde speziell für Anwendungen im medizinischen und ästhetischen Bereich von TBH entwickelt, um Reinheit, Infektionsschutz und Sauberkeit zu gewährleisten. Es wurde durch das unabhängige, akkreditierte Prüflabor CleanControlling Medical GmbH & Co. KG in Emmingen-Liptingen geprüft und bestätigt.

 

Um das TBH-Hygienekonzept umzusetzen, empfehlen wir folgende Schritte:

 

  1. Wechseln Sie den HEPA-Hauptfilter in der Anlage halbjährlich.

 

  1. Reinigen Sie die Absaughaube per Wischdesinfektion mit Desinfektionstüchern und reinigen Sie das Schutzgitter im Thermodesinfektor.

 

  1. Entnehmen Sie den InLine-Filter und setzen Sie einen neuen ein. Den gebrauchten InLine-Filter verpacken Sie sicher in einem geschlossenen, reißfesten Beutel und entsorgen ihn.

 

Zusätzlicher Tipp: Nur die effektive Erfassung und Filterung in Kombination mit der Reinigung und Desinfektion von Oberflächen schützt vor der Verschleppung von Viren/Bakterien zwischen Patienten. Lassen Sie die Anlage auch zwischen den Behandlungen eingeschaltet, damit sie mehrmals pro Stunde die gesamte Luft im Behandlungsraum filtern kann, wodurch das Kontaminations- und Infektionsrisiko deutlich reduziert wird.

Je nach Anwendungsfall und Hygienevorschriften in Ihrem Behandlungsraum kann es erforderlich sein, ein Hygienekonzept zu erstellen, das die Filter- und Absauganlage berücksichtigt.

 

In Bezug auf die Anlage sind insbesondere folgende Punkte wichtig und individuell zu bewerten:

 

– Empfehlung: Reinigung nach jeder Behandlung /// Mindeststandard: Reinigung pro Arbeitstag

  o Wischdesinfektion von Haube und Absaugarm (außen)

  o Wechsel des InLine-Filters

 

– Empfehlung: Reinigung wöchentlich /// Mindeststandard: Reinigung monatlich

  o Demontage und Reinigung des Absaugarms

  o Äußere Reinigung des Gerätes

 

– Empfehlung: halbjährliche Wartung /// Mindeststandard: jährliche Wartung

  o Durchführung der elektrischen Sicherheitsprüfung nach DGUV 3

  o Hauptfilter und Aktivkohlefilter wechseln

  o Äußere Reinigung der Anlage

Der InLine-Filter verhindert durch seine Position an der Spitze die Verschmutzung des Absaugarms, wodurch die Anhaftung von Partikeln und das Wachstum von Keimen und Bakterien reduziert wird. Der Aufwand für die Reinigung der Anlage wird dadurch erheblich reduziert. Der InLine-Filter ermöglicht es, dass die aufwändige Demontage und Reinigung des Absaugarms nicht mehr täglich, sondern nur noch wöchentlich oder monatlich durchgeführt werden muss, was Arbeitszeit spart und gleichzeitig die Hygiene sicherstellt.

 

– Darüber hinaus verhindert der InLine-Filter, dass Partikel und eventuell Flüssigkeiten aus dem Absaugarm herausfallen, wodurch eine Kreuzkontamination zwischen Patienten vermieden wird.

 

– Der InLine-Filter schützt zudem den Hauptfilter der Anlage, was dessen Lebensdauer erheblich verlängert.

 

– Die Absaugleistung der Anlage führt mehrmals pro Stunde zu einem kompletten Luftaustausch im Behandlungsraum, was zu einer deutlich geringeren Belastung der Luft mit Keimen und Bakterien führt.

 

– Die optimale Reinigungsfähigkeit wurde durch das unabhängige, akkreditierte Prüflabor CleanControlling Medical GmbH & Co. KG in Emmingen-Liptingen getestet und bestätigt.

 

Weitere Informationen finden Sie im Reinigungsvideo unter dem angegebenen Link: Luftabsaugung für medizinische Anwendungen – TBH GmbH  

No, the TBH filter and extraction systems are not medical devices according to § 3 para. 1 MPG, (German Medical Products Law, as they do not come into direct contact with patients and are primarily used for occupational safety. They do not meet the criteria for medical approval according to DIN EN 60601-1. However, they can be tested according to DIN EN 62353 (VDE 0751-1) to meet the increased safety requirements in the medical field.

No, clearance is not required because the systems are designed, and the extraction arm is positioned so that they do not directly touch the patient during extraction and application.

TBH Health Basic and TBH Health Pro have many features in common, such as similar air flow rates up to 200m³/h, identical dimensions, and identical filter types, including the patented InLine filter.

The main difference between the two units is the type of motor and the static pressure. The TBH Health Basic has a brush motor with a guaranteed lifetime of 600 operating hours and an average lifetime of approximately 600-800 hours. Although it is technically possible to change only the brushes, this can shorten the life of the motor, as the commutator is often also damaged, resulting in faster wear of the brushes. The TBH Health Basic offers a higher static pressure of 14,000 Pa, but this is irrelevant for extraction arm applications.

Due to the limited motor life and maintenance requirements, we recommend the TBH Health Basic for applications where the operating time is between 5 and 20 minutes per day. The TBH Health Pro for applications where the operating time exceeds 20 minutes per day.

Der Hauptunterschied zwischen den beiden Geräten liegt im Motortyp und im statischen Druck. Die TBH Health Basic verfügt über einen Bürstenmotor mit einer garantierten Laufzeit von 600 Betriebsstunden und einer durchschnittlichen Lebensdauer von ca. 600-800 Stunden. Obwohl es technisch möglich ist, nur die Bürsten zu wechseln, kann dies die Lebensdauer des Motors verkürzen, da häufig auch der Kollektor beschädigt wird, was zu einem schnelleren Verschleiß der Bürsten führt. Darüber hinaus bietet der TBH Health Basic einen höheren statischen Druck von 14.000 Pa, was jedoch für Anwendungen mit Absaugarmen irrelevant ist.

Aufgrund der begrenzten Lebensdauer des Motors und der Wartungsanforderungen empfehlen wir die TBH Health Basic für Anwendungen, bei denen die Betriebszeit zwischen 5 und 20 Minuten pro Tag liegt. Die TBH Health Pro für Anwendungen mit einer Betriebszeit von über 20min am Tag.

The TBH Health Pro and TBH Health Pro Silence are similar in technology and application, have similar airflow rates (max. 200m³/h or max. 300m³/h) due to the same motor technology and are maintenance free due to the brushless motor and drive system. The lifetime of the motor is approximately 20,000h – 40,000h and we give a warranty of 10,000h. Both systems are equipped with the patented InLine filter and are recommended for applications where the operating time is more than 20 minutes per day. However, there are significant differences: The TBH Health Pro is more compact, has a smaller filter area and a higher noise level (62 dB). The system uses a compact 2-stage filter (HEPA particle filter H13 + activated carbon). The TBH Health Pro Silence, on the other hand, requires slightly more space, but offers a larger filter area and a lower noise level (53 dB). It allows the HEPA filter H14 and the activated carbon filter to be replaced separately, resulting in lower filter replacement costs.

Changing the main filters in the TBH Health System is easy and does not require any special knowledge or tools. All you have to do is open the cover of the system, remove the old filters and insert the new ones. A detailed video tutorial can be found at Medical & Aesthetic Systems – TBH GmbH Extraction Systems: Medical & Aesthetic systems – TBH GmbH extraction systems

The assembly is very simple and does not require any special knowledge. Just watch our video tutorial at the link below: Medical & Aesthetic systems – TBH GmbH extraction systems

The extraction arm can be precisely positioned by turning the screws. Overstretching the extraction arm should be avoided as this can cause the system to tilt. See our video tutorial at the link below: Medical & Aesthetic systems – TBH GmbH extraction systems

Darüber hinaus sollten die Einstellschrauben wöchentlich nachgezogen werden, um ein mögliches Absinken des Absaugarms zu verhindern. Weitere Informationen finden Sie in den Anlagenkatalogen sowie in den Betriebs- und Wartungsanleitungen.

The TBH Hygiene Concept was developed specifically for TBH medical and aesthetic applications to ensure cleanliness and protection against infection. It has been tested by the independent, accredited test laboratory CleanControlling Medical GmbH & Co. KG in Emmingen-Liptingen, Germany.

To implement the TBH hygiene concept, we recommend the following steps:

  1. Change the main HEPA filter in the system every six months.
  2. Clean the extraction hood by wiping with disinfectant wipes and clean the protective grille in the thermal washer-disinfector.
  3. Remove the InLine filter and replace with a new InLine filter. Securely place the used InLine filter in a sealed, tear-proof bag for disposal.

Die Saugspitze hat einen kleineren Durchmesser. Dadurch wird die Sicht weniger behindert. Sie ist jedoch aus hygienischen Gründen nicht die erste Wahl, da sie den Absaugarm nicht vor Verschmutzung und Keimbildung schützt und den Hauptfilter stärker belastet. Sollte die serienmäßig montierte Haube bei der Positionierung am Patienten stören, kann sie durch Lösen der 4 Schrauben einfach abgenommen werden. Die Funktion des InLine-Filters wird dadurch nicht beeinträchtigt.

The TBH Health Pro is more compact and louder (62 dB), ideal for limited space and less continuous operation. The TBH Health Pro Silence requires more space, is quieter (53 dB) and has lower cost of ownership due to separately replaceable filters. It is better suited for continuous operation and noise sensitive environments. Both are designed for use for more than 20 minutes per day.

The TBH Health Series main filters are monitored by a differential pressure sensor that indicates when they are saturated. If saturation is not indicated within one year, it is necessary to change the HEPA particulate filters for sanitary reasons. The activated carbon filter, which is not monitored by a differential pressure sensor, should be changed at least every 6-12 months or when odors develop. According to the TBH hygiene concept, the InLine filter, which is also not monitored by a sensor, should be changed daily or weekly at the latest to prevent the transfer of bacteria and germs.

The system requires minimal maintenance. Other than regular filter changes and cleaning, the system is virtually maintenance free. The surfaces and the extraction arm should be cleaned regularly by wiping with disinfectant, see the step-by-step video instructions at: Medical & Aesthetic systems – TBH GmbH extraction systems

In addition, the adjustment screws should be tightened weekly to prevent possible sinking of the extraction arm. For more information, please refer to the equipment catalogs and operating and maintenance manuals.

The patented InLine filter is an innovation designed specifically for medical and aesthetic applications. Attached directly to the hood, the InLine filter captures particles, germs and bacteria at the point of origin. Thanks to the „1-click system“, it can be changed in about 5 seconds. The hood itself is easy to disassemble and clean. The InLine filter is located at the tip. This prevents the extraction arm from becoming dirty, where bacteria and germs can multiply and significantly reduces the amount of cleaning required for the system. In addition, the InLine filter prevents particles and possibly fluid from falling out of the suction arm, thus preventing the transfer of germs between patients. This protects the system’s main filter and significantly extends filter life.

The extraction tip is smaller in diameter, the view is less obstructed. However, it is not the first choice for hygienic reasons, as it does not protect the extraction arm from contamination and germ formation, and it puts more stress on the main filter. If the standard hood interferes with positioning on the patient, it can be easily removed by loosening the 4 screws. The function of the InLine filter is not affected.

The TBH main filters in the system are easy to change, no special knowledge or tools are required. Simply open the system’s cover, remove the filters and install the new ones. Click here for detailed video instructions: Medical & Aesthetic systems – TBH GmbH extraction systems

Changing the main filters in the TBH Health System is easy and does not require any special knowledge or tools. You just need to open the cover of the system, remove the old filters and insert the new ones. Click on the link for a detailed video tutorial: Medical & Aesthetic systems – TBH GmbH extraction systems

In Bezug auf die Anlage sind insbesondere folgende Punkte wichtig und individuell zu bewerten:

– Empfehlung: Reinigung nach jeder Behandlung /// Mindeststandard: Reinigung pro Arbeitstag

  o Wischdesinfektion von Haube und Absaugarm (außen)

  o Wechsel des InLine-Filters

– Empfehlung: Reinigung wöchentlich /// Mindeststandard: Reinigung monatlich

  o Demontage und Reinigung des Absaugarms

  o Äußere Reinigung des Gerätes

– Empfehlung: halbjährliche Wartung /// Mindeststandard: jährliche Wartung

  o Durchführung der elektrischen Sicherheitsprüfung nach DGUV 3

  o Hauptfilter und Aktivkohlefilter wechseln

  o Äußere Reinigung der Anlage

The assembly is very simple and does not require any special knowledge. Click on the link for a detailed video tutorial: Medical & Aesthetic systems – TBH GmbH extraction systems

– Darüber hinaus verhindert der InLine-Filter, dass Partikel und eventuell Flüssigkeiten aus dem Absaugarm herausfallen, wodurch eine Kreuzkontamination zwischen Patienten vermieden wird.

– Der InLine-Filter schützt zudem den Hauptfilter der Anlage, was dessen Lebensdauer erheblich verlängert.

– Die Absaugleistung der Anlage führt mehrmals pro Stunde zu einem kompletten Luftaustausch im Behandlungsraum, was zu einer deutlich geringeren Belastung der Luft mit Keimen und Bakterien führt.

– Die optimale Reinigungsfähigkeit wurde durch das unabhängige, akkreditierte Prüflabor CleanControlling Medical GmbH & Co. KG in Emmingen-Liptingen getestet und bestätigt.

Weitere Informationen finden Sie im Reinigungsvideo unter dem angegebenen Link: Luftabsaugung für medizinische Anwendungen – TBH GmbH  

The extraction arm can be precisely positioned by turning the screws. Overstretching the extraction arm should be avoided, otherwise the system may tilt. A detailed video tutorial can be found at the link: Medical & Aesthetic systems – TBH GmbH extraction systems

– Darüber hinaus verhindert der InLine-Filter, dass Partikel und eventuell Flüssigkeiten aus dem Absaugarm herausfallen, wodurch eine Kreuzkontamination zwischen Patienten vermieden wird.

– Der InLine-Filter schützt zudem den Hauptfilter der Anlage, was dessen Lebensdauer erheblich verlängert.

– Die Absaugleistung der Anlage führt mehrmals pro Stunde zu einem kompletten Luftaustausch im Behandlungsraum, was zu einer deutlich geringeren Belastung der Luft mit Keimen und Bakterien führt.

– Die optimale Reinigungsfähigkeit wurde durch das unabhängige, akkreditierte Prüflabor CleanControlling Medical GmbH & Co. KG in Emmingen-Liptingen getestet und bestätigt.

Weitere Informationen finden Sie im Reinigungsvideo unter dem angegebenen Link: Luftabsaugung für medizinische Anwendungen – TBH GmbH  

The TBH Hygiene Concept was developed specifically for TBH medical and aesthetic applications to ensure cleanliness, and protection against infection. It was tested by the independent, accredited test laboratory CleanControlling Medical GmbH & Co. KG in Emmingen-Liptingen, Germany.
To implement the TBH hygiene concept, we recommend the following steps:
1. Change the main HEPA filter in the system every six months.
2. Clean the suction hood by wiping it up with disinfectant wipes and clean the protective grille in the thermal disinfector.
3. Remove the InLine filter and replace it with a new one. Securely place the used InLine filter in a sealed, tear-proof bag for disposal.
Additional Tip: Only effective capture and filtration combined with surface cleaning and disinfection will prevent the spread of viruses/bacteria between patients. Leave the system on between treatments so it can filter all the air in the treatment room several times an hour, significantly reducing the risk of contamination and infection.

Depending on the application and the hygiene regulations in your treatment room, it may be necessary to create a hygiene concept that takes the filter and extraction system into account.

Regarding the system, the following points are important and should be evaluated individually:

– Recommendation: Cleaning after each treatment /// Minimum standard: Cleaning per working day.

  o Wipe disinfection of hood and extraction arm (external)

  o Change of InLine filter

– Recommendation: Cleaning weekly /// Minimum Standard: Cleaning monthly

  o Disassembly and cleaning of extraction arm

  o External cleaning of the unit

– Recommendation: Semi-annual maintenance /// Minimum Standard: Annual maintenance

o Perform electrical safety test according to DGUV 3

  o Change main filter and activated carbon filter

The InLine filter prevents contamination of the extraction arm due to its position at the tip, which reduces the adhesion of particles and the growth of germs and bacteria. This significantly reduces the effort required to clean the system. The InLine filter eliminates the need for a daily time-consuming disassembly and cleaning of the extraction arm, saving labor time while maintaining hygiene. Clean the extraction arm at least monthly, we recommend it weekly.

– In addition, the InLine filter prevents particles and liquids from falling out of the extraction arm, preventing cross-contamination between patients.

– The InLine filter also protects the system’s main filter, significantly extending its life.

– The system’s suction power creates a complete air exchange in the treatment room several times per hour, resulting in significantly less airborne germs and bacteria.

– The optimal cleaning performance was verified by the independent, accredited test laboratory CleanControlling Medical GmbH & Co. KG in Emmingen-Liptingen, Germany.

For more information, see the cleaning video at the link provided: Medical & Aesthetic systems – TBH GmbH extraction systems

No, i sistemi di aspirazione e filtraggio TBH non sono dispositivi medici ai sensi del § 3 par. 1 MPG, (legge tedesca sui dispositivi medici), in quanto non entrano in contatto diretto con i pazienti e sono utilizzati principalmente per la sicurezza sul lavoro. Non soddisfano i criteri per l’approvazione medica secondo la norma DIN EN 60601-1. Nel caso dovesse essere necessario, possono essere testati secondo la normativa DIN EN 62353 (VDE 0751-1) per soddisfare i maggiori requisiti di sicurezza in campo medicale.

No, non è obbligatorio ottenere una licenza classe lla perché il braccio di aspirazione è posizionato in modo tale da non toccare direttamente il paziente durante la fase di applicazione e aspirazione.

:  I dispositivi TBH Health Basic e TBH Health Pro hanno molti aspetti in comune, come la portata d’aria fino a 200m³/h, le dimensioni e gli stessi tipi di filtro, tra cui il filtro brevettato InLine.

La differenza principale tra le due apparecchiature è il tipo di motore e la pressione statica. La TBH Health Basic ha un motore a spazzole con un’autonomia garantita di 600 ore di funzionamento e una durata media di circa 600-800 ore. Sebbene sia tecnicamente possibile cambiare solo le spazzole, questo può ridurre la vita del motore, poiché spesso anche il collettore viene danneggiato, con conseguente usura più rapida delle spazzole. Il TBH Health Basic offre una pressione statica più elevata, pari a 14.000 Pa, che però è irrilevante per l’uso con bracci di aspirazione.

A causa della durata limitata del motore e dei requisiti di manutenzione, consigliamo il modello TBH Health Basic per le esigenze di funzionamento tra i 5 e i 20 minuti al giorno. Il TBH Health Pro per applicazioni con un tempo di funzionamento superiore a 20 minuti al giorno.

Il TBH Health Pro e il TBH Health Pro Silence sono simili per quanto riguarda la tecnologia e la loro applicazione, hanno portate d’aria simili (max. 200m³/h o max. 300m³/h) grazie alla medesima tecnologia del motore e sono esenti da manutenzione grazie al motore brushless senza spazzole (motore a commutazione elettrica azionato da una fonte di alimentazione in c.c. tramite un controllore esterno). La vita operativa del motore è di circa 20.000h – 40.000h, inoltre offriamo una garanzia di 10.000h. Entrambe le apparecchiature sono dotate del filtro brevettato InLine e sono indicate per esigenze di lavoro che richiedono più di 20 minuti al giorno. Ci sono però delle differenze importanti. Il TBH Health Pro è più compatto, ha un’area di filtraggio più piccola e un livello di rumorosità più elevato (62 dB). Il sistema utilizza un filtro compatto a 2 stadi (filtro HEPA per particelle H13 + carbone attivo). Il TBH Health Pro Silence, invece, richiede uno spazio leggermente maggiore, offre un’area di filtraggio più grande e un livello di rumorosità più basso (53 dB). Il filtro antiparticolato HEPA H14 e il filtro a carbone attivo possono essere sostituiti separatamente, con conseguente riduzione dei costi di manutenzione.

Il TBH Health Pro è più compatto e più rumoroso (62 dB), ideale per spazi limitati e per un utilizzo meno frequente. Il TBH Health Pro Silence richiede più spazio, è più silenzioso (53 dB) e ha costi di manutenzione inferiori grazie ai filtri che si possono sostituire separatamente. È più adatto al funzionamento continuo e agli ambienti più sensibili al rumore. Entrambi sono progettati per un utilizzo superiore ai 20 minuti al giorno.

I filtri principali della serie TBH Health sono monitorati da un sensore di pressione differenziale che indica quando sono saturi. Se la saturazione non viene segnalata entro un anno, è necessario cambiare i filtri HEPA per motivi igienici. Il filtro a carbone attivo, che non è monitorato da un sensore di pressione differenziale, deve essere sostituito almeno ogni 6-12 mesi o quando si sviluppano odori. Secondo il concetto di igiene di TBH, il filtro InLine, anch’esso non monitorato da un sensore, dovrebbe essere cambiato quotidianamente, al massimo settimanalmente, per evitare il contagio con batteri e germi.

La manutenzione del sistema è minima. A parte la sostituzione e la pulizia periodica dei filtri, il sistema è praticamente esente da manutenzione. Le superfici e il braccio di aspirazione devono essere puliti regolarmente con un panno disinfettante, vedi le istruzioni video step by step:

Aspiratori settore estetica e medicale – TBH abbattimento particelle

Inoltre, le viti di regolazione devono essere riavvitate settimanalmente per evitare che il braccio di aspirazione possa abbassarsi. Per ulteriori informazioni, consulta i cataloghi del sistema e le istruzioni per l’uso e la manutenzione.

: Il filtro InLine brevettato è un’innovazione pensata appositamente per le applicazioni medicali ed estetiche. Questo filtro, collegato direttamente alla parabola, abbatte particelle, germi e batteri direttamente al punto di origine. Grazie al „sistema 1-click“, il filtro InLine può essere sostituito in circa 5 secondi. Essendo posizionato sulla sommità, evita che il braccio di aspirazione si riempia di impurità, dove altrimenti potrebbero moltiplicarsi batteri e germi. La pulizia del sistema si riduce così in modo considerevole. Inoltre, il filtro InLine impedisce la fuoriuscita di particelle e di liquidi dal braccio di aspirazione e di conseguenza il passaggio di germi tra pazienti. Il filtro principale dell’unità è viene protetto e la sua durata di vita si prolunga notevolmente.

La punta di aspirazione ha un diametro più piccolo. Ciò significa che il campo visivo è meno limitato. Non è però la prima scelta per motivi igienici, in quanto non protegge il braccio di aspirazione dalla contaminazione e dalla formazione di germi e mette a dura prova il filtro principale. Se la parabola interferisce il campo visivo durante l’uso, può essere facilmente rimossa allentando le 4 viti.  L’uso senza parabola non influisce sul funzionamento del filtro InLine.

I filtri principali dei sistemi TBH sono facili da sostituire, non sono necessarie conoscenze o strumenti particolari. Basta aprire il coperchio del sistema, rimuovere i filtri e inserire quelli nuovi.

Consulta le istruzioni dettagliate video su: Aspiratori settore estetica e medicale – TBH abbattimento particelle

La sostituzione dei filtri principali dell’unità TBH Health è semplice e non richiede conoscenze o strumenti particolari. Tutto ciò che devi fare è aprire il coperchio dell’unità, rimuovere i filtri usati e inserire quelli nuovi. Puoi trovare un video dettagliato di istruzioni al link: Aspirazione dell’aria per applicazioni mediche – TBH GmbH

Il processo di montaggio è molto semplice e non richiede conoscenze particolari. Consulta le istruzioni dettagliate video su: Aspiratori settore estetica e medicale – TBH abbattimento particelle

Il braccio di aspirazione può essere posizionato con esattezza ruotando le viti. Si consiglia di non allungare eccessivamente il braccio di aspirazione, per evitare che l’apparecchio si possa sbilanciare e cadere. Consulta le istruzioni dettagliate video su: Aspiratori settore estetica e medicale – TBH abbattimento particelle

Il sistema di igiene TBH è stato sviluppato appositamente dalla TBH per applicazioni in campo medicale ed estetico, al fine di garantire sicurezza, protezione contro le contaminazioni e pulizia. È stato testato dal laboratorio indipendente e accreditato CleanControlling Medical GmbH & Co. KG di Emmingen-Liptingen in Germania.

Per implementare il piano igienico TBH, ti consigliamo di seguire i seguenti step:

1° Sostituire il filtro principale HEPA dell’unità ogni sei mesi.

2° Pulire il sistema di aspirazione mediante disinfezione con salviette disinfettanti e pulire la griglia di protezione nel termodisinfettore.

3° Estrarre il filtro InLine e inserirne uno nuovo. Porre il filtro InLine usato in un sacchetto antistrappo, e successivamente smaltirlo.

Suggerimento aggiuntivo: solo l’abbattimento, la pulizia e alla disinfezione delle superfici protegge contro la diffusione di virus/batteri tra i pazienti. Mantieni il dispositivo acceso tra un trattamento e l’altro in modo che possa filtrare più volte all’ora l’aria nell’ambiente, riducendo in modo significativo il rischio di contaminazione e infezione.

A seconda del tipo di applicazione e delle norme igieniche della tua sala di trattamento, potrebbe essere necessario creare un piano di igiene che prenda in considerazione il sistema di filtraggio e di aspirazione.

Per quanto riguarda il sistema, i punti seguenti sono particolarmente importanti e devono essere valutati individualmente:

– Raccomandazione: pulizia dopo ogni trattamento /// Standard minimo: pulizia ogni giorno lavorativo.

  o Disinfezione con panno della parabola e del braccio di aspirazione (all’esterno)

  o Sostituzione del filtro InLine

– Raccomandazione: pulizia settimanale /// Standard minimo: pulizia mensile

  o Smontaggio e pulizia del braccio di aspirazione

o Pulizia esterna dell’apparecchio

– Raccomandazione: manutenzione semestrale /// Standard minimo: manutenzione annuale

o Esecuzione dell’ispezione di sicurezza elettrica secondo la normativa DGUV 3

o Sostituzione del filtro principale e del filtro a carbone attivo

o Pulizia esterna dell’apparecchio

Il filtro InLine previene la contaminazione del braccio di aspirazione grazie alla sua posizione sulla sommità, riducendo così la presenza di particelle e la proliferazione di germi e batteri. L’impegno richiesto per la pulizia del sistema è quindi notevolmente ridotto. Il filtro InLine fa sì che le lunghe operazioni di smontaggio e pulizia del braccio di aspirazione non debbano più essere effettuate quotidianamente, ma solo settimanalmente o mensilmente, risparmiando tempo di lavoro e garantendo allo stesso tempo l’igiene.

– Inoltre, il filtro InLine impedisce la fuoriuscita di particelle e liquidi dal braccio di aspirazione, evitando così la contaminazione tra pazienti.

– Il filtro InLine protegge anche il filtro principale dell’unità, prolungandone notevolmente la durata.

– Le prestazioni di aspirazione dell’unità si concretizzano in un ricambio dell’aria nella sala di trattamento più volte all’ora, con una conseguente riduzione della contaminazione dell’aria.

– La funzionalità di pulizia eccellente è stata testata dal laboratorio di prova indipendente e accreditato CleanControlling Medical GmbH & Co. KG di Emmingen-Liptingen in Germania

Per maggiori informazioni, consulta il video sulla pulizia su: Aspiratori settore estetica e medicale – TBH abbattimento particelle